tbf/TBF/Rig/TestMethods/FlyingStart
Michal Buzik 52bd0d01ce Restore Genesis communication and Q3 calibration from special branches
A – Registration and execution
- Register GenesisCommunication Factory in the component list.
- Separate the Genesis form and sequence from iPerl communication.

B – Communication activities
- Restore initialization, connection, PCB reading, password and login.
- Include grouped login, mode switching and disconnection.
- Support processing up to 10 slots.

C1 – Input calibration factors
- Read three factors from Water meters / Text1–Text3.
- Validate integer values in the range 1–65535.
- Preserve the default of 15625 when all three fields are empty.

D – Q3 calibration
- Connect the Prepare Q3 → measurement → Write Q3 workflow.
- Add channel processing to FlyingStart and FlyingStartMassCollection.
- Reset previous measurement data and validate calculated factors.
- Mark factors as stored only after StoreCalibration succeeds.

E – Results and database
- Store calibration factors separately for each meter and channel.
- Add result entities, mappings and Q3 data.
- Extend DB.cs / EnsureSchema to create and update the schema.
- Preserve compatibility with the existing binary format.

Validation:
- Debug build and 18 tests passed.
- Simulated communication runs follow the same activity sequence.
- The complete Q3 workflow has not yet been verified on hardware.

Known limitation:
- An inherited mismatch in simulated responses and error propagation
  can produce an incorrect OK result; this change does not fix it.
2026-09-09 15:04:49 +02:00
..
Compound iPerl ASIC - implementation 2026-03-12 14:50:47 +01:00
HeatMeters iPerl ASIC - implementation 2026-03-12 14:50:47 +01:00
Single iPerl ASIC - implementation 2026-03-12 14:50:47 +01:00
FlyingStartSeq.cs Restore Genesis communication and Q3 calibration from special branches 2026-09-09 15:04:49 +02:00